Product Description
Description
Free chlorine exists in solution as pH dependent ratio of hypochlorous acid (~100% at pH 5) and hypochlorite ion (~100% at pH 10). Free Chlorine is the sum of the Hypochlorous acid and Hypochlorite ion in the sample. The Free Chlorine Sensor measures only the hypochlorous acid component of the free chlorine and the analyzer calculates the balance using either the measured or a user defined fixed value. The use of the pH sensor provides accurate compensation for samples between pH 6 and pH 9.5 and eliminates the need for an expensive sample conditioning system to control the pH of the solution.
SC9184_FCO Sensor consist of the parts below
- Sensor flow cell (PMMA)
- Chlorine probe (non-reagent type) with glass beads cleaning.
- PH sensor for PH compensation.
- Flow control valve
- Optional flow measuring sensor
- Electronics unit with RS485 digital output by Modbus RTU protocol Note
Specification
Range | (0.05~5.00/10.00)mg/l, ppm |
Accuracy | 0.01mg/l (0~0.5)mg/l, ppm 0.02mg/l (0.5~2)mg/l, ppm 0.1mg/l (2~5)mg/l, ppm 0.2mg/l (5~10)mg/l, ppm |
Resolution | 0.01mg/l |
Repeatability | ±1%FS |
Auto Temp. compensation | 0- 65°C |
pH compensation | 5-9pH |
Electrodes | Cathode Pt, Anode Cu |
Sample flow rate | 2.5l/min |
Operate Pressure | 1- 14.7bar |
Operate Temp. | 0- 65°C |
Response Time | T90<3s |
Water Proof | IP68 |
Temp. Sensor | Pt100 |
Temp. Range | 0~65°C |
Temp. Accuracy | ±1°C |
Output | RS485 Modbus RTU |
Material of housing | PP |
-
- GDC-01/02 Transmitter GDC-04/06/08 Controller GDC-H Transmitter
-
- Single or dual-channels Multi-channels up to 8 Handheld Terminal