Cyanobacteria Sensor

Description

SC9260-BA Cyanobacteria (Blue-green Algae) sensor uses fluorescence approach to measure cyanobacteria concentrations in water. When cyanobacteria are excited by a high-energy LED light source of 590nm, it produces a specific wavelength fluorescence which based on its spectrum absorption characteristics, can be used to detect cyanobacteria concentrations in water. SC9260-BA sensor is particularly suitable for scientific research as well as for online water quality monitoring of ground water, surface water and other water applications.

Specification

Measuring system Light Source: LED (590 nm)
Detector: Photo diode + filter (655 nm)
Principle Fluorescence
Measuring range 0 to 20 ppb, 500 ppb
Accuracy 0.02 ppb
Respond Time T90 < 10s
Measuring Interval 5 s
Operate Pressure 3 bar, 1 bar in Flow cell 2 to 4 L/min
Operate Temp. 32 to 104 °F (0 to 40 °C)
Power 12 to 30VDC, Max. 3.5W
Digital Interface RS485 Modbus RTU
Housing Material: 316SS, Titanium is Optional
Auto Cleaning Optional Clamp-on Wiper
Protection Rate IP68, submersible
Dimension Dia. 2” (50.8 mm), Length 8.66” (220 mm)
Weight 6 lbs (2.8 kg)

Feature

  • High sensitivity, fast response, stable and reliable
  • Range adjustable, automatic daylight compensation
  • Low power consumption, easy operation and maintenance
  • Digital transmission, anti-interference ability
  • Integrated sensor design, submersible to operate down to 300 meters.

Application

  • Lake, reservoir, underground and river water
    Hydrocarbons are dangerous pollutants that have to be detected very early on reservoir or river for producing drinking water. Fast and reliable measuring system is necessary, and the UV fluorescence is the only method matching these requirements.
  • Cooling water survey
    The concentration of hydrocarbons in cooling water is a critical parameter on refineries.
  • Wastewater monitoring
    The effluents of refineries and chemical plants have to stay below limits to meet the environmental regulations. A reliable and low maintenance measuring system is required and only UV fluorescence approach achieves this goal.